Year : 2010  |  Volume : 3  |  Issue : 2  |  Page : 182-184
Improper tube fixation causing a leaky cuff


Department of Anesthesia and Intensive Care, Jai Prakash Narayan Apex Trauma Centre, All India Institute of Medical Sciences, Ansari Nagar, New Delhi-110 029, India

Click here for correspondence address and email

Date of Submission11-Sep-2008
Date of Acceptance24-Mar-2009
Date of Web Publication31-Mar-2010
 

   Abstract 

Leaking endotracheal tube cuffs are common problems in intensive care units. We report a case wherein the inflation tube was damaged by the adhesive plaster used for tube fixation and resulted in leaking endotracheal tube cuff. We also give some suggestions regarding the tube fixation and some remedial measures for damaged inflation system.

Keywords: Improper fixation, leaky cuff, inflation system

How to cite this article:
Gupta B, Farooque K, Jain D, Kapoor R. Improper tube fixation causing a leaky cuff. J Emerg Trauma Shock 2010;3:182-4

How to cite this URL:
Gupta B, Farooque K, Jain D, Kapoor R. Improper tube fixation causing a leaky cuff. J Emerg Trauma Shock [serial online] 2010 [cited 2019 Nov 13];3:182-4. Available from: http://www.onlinejets.org/text.asp?2010/3/2/182/62125



   Introduction Top


A well-secured endotracheal tube (ETT) is essential for safe anesthesia. If ETT is not well secured, there is a danger of either unplanned extubation or advancement of the tube deeper into the trachea. However, improper tube fixation can damage the inflation system and can be a rare cause of leaky ETT cuff, making maintenance of adequate ventilation difficult. We encountered a case wherein the inflation tube was damaged by the adhesive plaster used for tube fixation.


   Case Report Top


A 33-year-old male patient with the history of motor vehicular accident was brought in the emergency department. On primary survey, his airway was compromised and oxygen saturation with O 2 by face mask was 78%. His Glasgow goma score was 5; hence to protect the airway, the patient was intubated with 8.0 mm ETT by a resident doctor. The tube was secured with Durapore adhesive plaster at 21 cm at the level of incisors. The adhesive plaster was encircled thrice around the tube; once including the tube and twice excluding the inflation tube. The patient was further investigated and CT scan head revealed a large extradural hematoma with midline shift. The patient was posted for surgical evacuation of extradural hematoma. In operation theatre, it was observed that the ETT cuff was leaking. After checking the exposed part of inflation system, which was intact, it was decided to change the tube. On removing the adhesive plaster, we observed that there was a cut in the inflation tube at the level of entry in ETT. The trachea was reintubated with 8.0 mm ETT. The difficult airway kit, that is, multiple blades, small sized tubes, McCoy laryngoscope, laryngeal mask airway, and tracheostomy kit were kept ready in case of any difficulty in reintubation. On close examination of ETT, we concluded that overextension of the inflation tube lead to the cut in the tubing. Three turns of adhesive plaster once including and twice excluding the inflation tube might be responsible for avulsion of the inflation tube from the tracheal tube. Rest of surgery was uneventful. The patient was on ventilator for 4 days. He was discharged on day 12 after surgery with mild cognitive dysfunction.


   Discussion Top


A leaking endotracheal cuff may make maintenance of adequate ventilation difficult, fail to protect against aspiration, and make surgery difficult. A defect in the cuff or the inflation system, i.e., inflation tube, pilot balloon, or the valve may be the reasons of leaking cuff. When one is faced with this problem, the best solution is to have the patient reintubated as soon as possible. This is especially true if the patient is mechanically ventilated, because little tidal volume will be delivered while the cuff is unpressurised and deflated. However, extubation, manual ventilation, and reintubation should be avoided, if possible, in certain situations [1] :

  • When a mechanically ventilated patient is receiving a high oxygen percentage and/or PEEP level
  • Anticipated difficult intubation-it is very important to assess the airway and identify difficult intubation to avoid any catastrophe. The LEMON scale is one of the available predictive methods to assess difficult intubation. [2] The score with a maximum of 10 points is calculated by assigning 1 point for each of the following


LEMON criteria

  • L = Look externally for any characteristics that are known to cause difficult intubation (facial trauma, large incisors, beard or moustache, large tongue)
  • E = Evaluate the 3-3-2 rule (incisor distance, 3 finger breadths; hyoid-mental distance, 3 finger breadths; thyroid to floor of mouth distance, 2 finger breadths)
  • M = Mallampati classification; used to visualize the hypopharynx. [3],[4] It is not very useful in emergency situations
  • O = Obstruction which makes laryngoscopy and intubation difficult (presence of any condition like epiglottitis , peritonsillar abscess, trauma)
  • N = Neck mobility (limited neck mobility)
Patients in the difficult intubation group have higher LEMON scores. [2],[5]

  • When a patient has had upper airway trauma or recent surgery
  • When vomiting and aspiration are likely
  • When patient has had a recent tracheostomy
  • When trained personnel or intubation equipment are not available


We would like to give some suggestions regarding tube fixation and some temporary remedial measures in case of leaky cuff if reintubation is not feasible:

  1. The above incident reprimands everybody that correct basic techniques should not be ignored as they can lead to disasters in case of difficult intubation.
  2. While fixing the tube with adhesive plaster, it should not be overlapped multiple times, as it can lead to leaking cuff as in our case and also make its removal difficult during extubation [Figure 1].
  3. The basic principle of traction and countertraction should be followed while securing the tube [Figure 2].
  4. Even in case of simple intubations, junior doctors should be supervised and trained for securing the tube.
  5. In case of difficult intubation and if the inflation tube is cut at the level of entry into tracheal tube, the technique described by Sprung Juraj [6] may be used. He used the inflation tube of an unused ETT, and joined to remnant section of the inflation tube on the patient's ETT using a 0.5-1-inch section of a standard hypodermic needle [Figure 3]. [6] Fisher also described a similar technique wherein cut ends of the inflation tube were pulled together over the needle. The inflation tube was sprayed with tincture benzoin and wrapped with plaster. [7] In case of cracked one-way valve or cut pilot balloon or inflating tube, Jims Sill described a technique that can be used to reinflate the cuff. [1] The assembled device consists of a 10-ml syringe, a three-way stopcock, and a needle. The needle is inserted into the severed inflating tube. Opening the stopcock to only the syringe and needle permits the cuff to be inflated without the monitoring of pressure [Figure 4]. Watson also gave some simple solutions for leaking ETT cuffs due to incompetent pilot balloon valve. [8] T connector can be connected to the pilot balloon to function as a secondary valve for the system [Figure 5]. The plastic clip of the "T" connector can be loosened to add air to the system and then clamped to keep it pressurized. If the cuff still leaks, the problem may be a hole in the pilot balloon. If this is suspected, the pilot balloon is cut from ETT and a 22-guage intravenous catheter that has been moistened with alcohol is threaded directly into the lumen of inflation tube. Alcohol helps lubricate the catheter and then cements it in the tube after evaporating. The other end of the catheter may be attached to either stopcock [Figure 6] or T connector to keep the cuff pressurized.



   Conclusion Top


We have reported a case of leaking endotracheal cuff due to faulty fixation of tube. Securing the tube is a basic maneuver, yet if done improperly can lead to disasters. The basic principle of traction and countertraction should be followed for ETT fixation and multiple overlaps of adhesive plaster should be avoided. In case of difficult intubation or situations wherein reintubation is not possible, simple techniques described earlier can be used to reinflate the cuff.

 
   References Top

1.Sills J. An emergency cuff inflation technique. Respir Care 1986;31:199-201.   Back to cited text no. 1      
2.Mallampati SR, Gatt SP, Gugino LD, Waraksa B, Freiburger D, Liu PL. A Clinical sign to predict difficult intubation: A prospective study. Can Anaesth Soc J 1985;32:429-34.   Back to cited text no. 2      
3.Samsoon GL, Young JR. Difficult tracheal intubation: A retrospective study. Anaesthesia 1987;42:487-90.   Back to cited text no. 3  [PUBMED]    
4.Murphy MF, Walls RM. The difficult and failed airway. Manual of emergency airway management. Chicago: Lipincott Williams and Wilkins; 2000. p. 31-9.   Back to cited text no. 4      
5.Reed MJ, Dunn MJ, McKeown DW. Can an airway assessment score predict difficulty at intubation in the emergency department? Emerg Med J 2005;22:99-102.  Back to cited text no. 5  [PUBMED]  [FULLTEXT]  
6.Sprung J, Bourke DL, Thomas P, Harrison C. Clever cure for an endotracheal cuff leak. Anesthesthesiology 1994;81:790-1.  Back to cited text no. 6      
7.Fisher MM. Repairing pilot balloon lines. Anaesth Intens Care 1988;16:501.  Back to cited text no. 7      
8.Watson E, Harris MM. Leaking endotracheal tube. Chest 1989;95:709.   Back to cited text no. 8  [PUBMED]    

Top
Correspondence Address:
Babita Gupta
Department of Anesthesia and Intensive Care, Jai Prakash Narayan Apex Trauma Centre, All India Institute of Medical Sciences, Ansari Nagar, New Delhi-110 029
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0974-2700.62125

Rights and Permissions


    Figures

  [Figure 1], [Figure 2], [Figure 3], [Figure 4], [Figure 5], [Figure 6]

This article has been cited by
1 Korrelliert bei schwerverletzten Patienten der Durchmesser der V. cava inferior im CT mit der Sterblichkeit?
E.C. Schubert,S. Huber-Wagner,K. Holzapfel
Notfall + Rettungsmedizin. 2014;
[Pubmed] | [DOI]
2 Initial inferior vena cava diameter on computed tomographic scan independently predicts mortality in severely injured trauma patients
Jeremy J. Johnson,Tabitha Garwe,Roxie M. Albrecht,Ademola Adeseye,David Bishop,Robert B. Fails,David W. Shepherd,Jason S. Lees
Journal of Trauma and Acute Care Surgery. 2013; 74(3): 741
[Pubmed] | [DOI]
3 Praxis der Fixierung des Endotrachealtubus
O. Rothaug,A. Müller-Wolff,R. Kaltwasser,R. Dubb,C. Hermes
Medizinische Klinik - Intensivmedizin und Notfallmedizin. 2013; 108(6): 507
[Pubmed] | [DOI]
4 Endotracheal Tube Cuff Leaks
Mohammad El-Orbany,M. Ramez Salem
Anesthesia & Analgesia. 2013; 117(2): 428
[Pubmed] | [DOI]
5 Endotracheal tube cuff leaks: Causes, consequences, and management
El-Orbany, M. and Salem, M.R.
Anesthesia and Analgesia. 2013; 117(2): 428-434
[Pubmed]
6 Failure to deflate endotracheal tube cuff due to improper fixation
Jain, D. and Bala, I.
Sri Lankan Journal of Anaesthesiology. 2013; 21(2): 90
[Pubmed]
7 Incorrect fixation of endotracheal tube: A cause of non-inflation of the cuff
Jangra, K. and Malhotra, S.K.
Middle East Journal of Anesthesiology. 2012; 21(6): 869-870
[Pubmed]
8 Accidental intraoperative avulsion of external inflation tubing of armored endotracheal tube
Bhandari, S. and Gupta, S.P. and Gupta, K. and Kumar, A.
Journal of Anaesthesiology Clinical Pharmacology. 2012; 28(1): 132-133
[Pubmed]
9 Reply to Letter to the Editor “Most Suitable Modality to Assess the IVC in the Trauma Patient”
Yu-Ying Liao,Kuo-Tai Chen
The Journal of Trauma: Injury, Infection, and Critical Care. 2011; 71(6): 1925
[Pubmed] | [DOI]



 

Top
 
  Search
 
  
  
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Email Alert *
    Add to My List *
* Registration required (free)  


    Abstract
    Introduction
    Case Report
    Discussion
    Conclusion
    References
    Article Figures

 Article Access Statistics
    Viewed2850    
    Printed174    
    Emailed0    
    PDF Downloaded33    
    Comments [Add]    
    Cited by others 9    

Recommend this journal